Wheatstone Bridge Calculator: Resistance or Voltage
The Wheatstone bridge compares two voltage dividers wired in parallel across the same supply, to measure a resistance more precisely than a plain ohmmeter can — it's the classic building block behind strain gauges and resistive sensors (temperature, pressure, strain). A galvanometer wired between the two midpoints B and D reads the difference: zero once the bridge is balanced.
Enter the 4 resistors and the supply voltage to get the unbalance voltage Vg = VB − VD; or, if you're after the resistor that exactly balances the bridge, enter R1, R2, R3 and let the page work out R4.
The bridge
Why a Wheatstone bridge measures more precisely than a plain ohmmeter
A cheap ohmmeter forces a current and reads a drop. That reading moves with the supply, the lead resistance, and however honest the current source felt that morning. A Wheatstone bridge is a comparison. You null the voltage between two mid-points. At null, only the ratio of resistors matters. The excitation can wander; the null stays put.
That is why a 19th-century galvanometer and four resistors still beat a 3½-digit meter on a small change of resistance.
Proof of the balance condition R1·R4 = R2·R3
Two voltage dividers on the same excitation Vs: R1–R2 on one side, R3–R4 on the other (numbering as on this page). The galvanometer sits between the taps.
VB = Vs · R2 / (R1 + R2) · VD = Vs · R4 / (R3 + R4)
Null means VB = VD. Cancel Vs, cross-multiply:
R1 · R4 = R2 · R3 ⟺ Vg = 0
Know three, find the fourth. Off-null, the page's difference of the two divider formulae gives Vg directly.
Strain gauges: full bridge, half bridge, quarter bridge
A foil gauge changes a few milliohms per microstrain. The bridge is how you see it.
- Quarter bridge — one active gauge, three fixed (or a dummy). Least sensitive; temperature of that one gauge walks straight into the reading.
- Half bridge — two actives, often tension and compression on a beam. Temperature cancels, signal doubles.
- Full bridge — four actives. Maximum output, best compensation, the usual load-cell arrangement.
A “dummy” gauge, bonded to unstressed material next to the active one, is the poor man's half bridge for temperature only.
Amplifying the unbalance voltage: instrumentation amplifier setup
Vg is millivolts, sometimes microvolts, sitting on a common-mode of Vs/2. You want a high-impedance, high-CMRR amplifier — an instrumentation amp, gain set by one resistor.
Do not load the bridge with a 10 kΩ op-amp inverter. The divider ratio moves. An in-amp (or a discrete three-op-amp in-amp) is the circuit this measurement has used for fifty years because the alternatives cheat.
The AC Wheatstone bridge: impedance measurement
Swap resistors for impedances, drive with AC, and balance both magnitude and phase:
Z1 · Z4 = Z2 · Z3
Maxwell, Wien, Schering — named bridges are just choices of which arm is the C or the L. The detector is a scope or a tuned voltmeter, not a DC galvanometer. Same null idea.
Bridge sensitivity depending on where the variable resistor sits
dVg / (dR/R) is largest when the four arms are equal, or at least the same order. A 1 kΩ gauge in a 100 kΩ opposite arm is a deaf bridge. Put the rheostat in the arm that keeps the ratios near 1, and trim so that the working range sits around null, where the response is almost linear.
Applications: pressure sensors, RTD temperature sensors, strain sensors
- Pressure — a diaphragm with gauges in a full bridge. The millivolt-per-volt figure on the datasheet is this Vg / Vs.
- RTD (Pt100) — platinum in one arm. Three-wire or four-wire connections exist to kill lead resistance; the bridge (or a constant-current cousin) is still the front end.
- Strain — beams, bolts, airframes. Gauge factor × strain = ΔR/R, which the bridge turns into Vg.
Measurement errors: lead resistance, self-heating
- Leads — 0.1 Ω of copper on a 120 Ω gauge is already 800 µε of fake strain. Four-wire Kelvin to the gauge, or a three-wire RTD arrangement, is not optional on a long cable.
- Self-heating — excitation current warms the element. Pt100 is often run at 0.5–1 mA so the I²R rise stays a fraction of a degree. More volts, more signal, more lie.
- Tempco of the other arms — metal-film, matched, or the “fixed” arms become a second thermometer.
Gauge factor and the millivolt-per-volt figure
Gauge factor GF ≈ 2 for constantan foil: ΔR/R = GF · ε. A 1000 µε on a 350 Ω gauge is 0.70 Ω. In a 5 V quarter bridge that is a couple of millivolts at the diagonal — hence “mV/V” on load-cell labels. Double or quadruple that for half and full bridges with gauges in the right signs.
Three-wire and four-wire RTD connections
Two wires to a Pt100 add the copper twice into the arm. Three-wire puts equal leads in adjacent arms so they cancel at balance. Four-wire abandons the bridge for a force/sense current pair. Both are legal; mixing them in one junction box is how people invent a 2 °C offset.
Why the galvanometer current is not the reading
Off-null, a real galvanometer (finite Rm) loads the Thevenin of the two dividers and Vg sags. The formula on this page is the open-circuit diagonal voltage — what a high-impedance voltmeter or in-amp sees. A 50 Ω meter from a 19th-century kit will not match the number, and that is expected.
Balance (Vg = 0) is independent of Vs. The off-null voltage is not: it scales with excitation. Quote mV/V, or quote both the millivolts and the supply, never the millivolts alone.